Q-1300A Aerosol generator is a special instrument for generating DOP aerosol by Laskin nozzle. The embedded adjusting valve could control the working of 4 or 10 nozzles. When the air flow is 1.4m3/min-56.6m3/min, the output aerosol concentration is 10μg/L-100μg/L. The performance of aerosol complies to national standards, and the device is suitable to institutes of medical device testing, centers of disease control, hospitals, pharmaceutical companies, HEPA filter manufacturers for leakage testing of clean rooms and HEPA filters.
Standards
GB 50591-2010 Code for construction and acceptance of cleanroom
YY0569-2005 Biosafety cabinet
GB/T13554-2008 High efficiency particulate air filter
Features
Unique air channel design, stable air flow and balanced particle output. Generate multiple kinds of aerosols, DOP, DOS, PAO….
Nebulizing concentration is adjustable in a wide range.
The Main Parameters | Parameter Range | Resolution | Maximum Permissible Error (MPE) |
Working pressure | (60-150) kPa | 1 kPa | ±0.5% |
Particle output range | (1.4-56.6) m3/min | ||
Particle concentration | 100μg/L (air flow is 5.6 m3/min) | ||
Particle concentration | 10μg/L(air flow is 56.6 m3/min) | ||
Generating method | 4-10 Laskin nozzle | ||
Compressed air | Embedded compressor | ||
Air type | Multiple size of particles (Cold generating) | ||
Dimension | (Length 200×Width 500×Height 280) mm | ||
Noise | <65dB(A) | ||
Weight | About 18Kg | ||
Working power | AC220V±10%,50Hz | ||
Power consumption | ≤500W |