Measuring Range:
High Frequency Induction Furnace(500mg Sample): LC: 1ppm - 0.1% HC: 0.1%~5% LS: 1ppm - 0.3% HS: 150mg Sample): 0.3%~30%
Tubular Resistance Furnace(500mg Sample): C: 0.01% ~20% S: 0.005% ~20%*
Note: * the measurement range can be changed by changing the measurement quantity. *The number of detection cells and sizes can be determined in accordance with the content of carbon and sulfur in specimens of users
Sensitivity:
Carbon and Sulfur: 0.01ppm
Accuracy: High Frequency Induction Furnace: LC: 1ppm or RSD≤1% HC: RSD≤0.5% LS: 1.5ppm or RSD≤1.5% HS: RSD≤1.5%
Tubular Resistance Furnace: LC: RSD≤1% HC: RSD≤0.5% LS: RSD≤1.5% HS: RSD≤1.5%
Analysis Time:
High Frequency Induction Furnace: 40-60 seconds
Tubular Resistance Furnace: 150-300 seconds
Specimen Weighting:
High Frequency Induction Furnace: 0.1g-0.5g
Tubular Resistance Furnace: 300mg/coal(typical value)
Combustion Furnace:
High Frequency Induction Furnace: 18MHz 2.7KVA
Tubular Resistance Furnace: Max.20A,Max.1550℃, the temperature can be adjusted continuously
Dust Sweeping:
The dust in the combustion tube of the high frequency induction furnace is wiped off automatically
Carrier Gas:
Oxygen 99.5%, 3L/min
Power Gas: Nitrogen or Compressed Air
Chemical Reagent:
Magnesium perchlorate,CO2 absorbent
Power Supply:
Power Supply of the Analyzer: 220VAC ±10%,50±1Hz,16A
Power Supply of the Resistance Furnace: 220VAC ±10% , 50±1Hz, 20A
Weight:
High Frequency Induction Furnace ~100kg,
Tubular Resistance Furnace ~30kg
Size:
The Analyzer: 55x77x76cm(W×H×D)
Resistance Furnace: 33×52×60cm(W×H×D)
Resolution of the Electronic Balance: 0.1mg