Determination of azide in irbesartan by ion chromatography

2020-08-15 00:141004
Return

Irbesartan

Irbesartan is an angiotensin Ⅱ (Ang Ⅱ) receptor inhibitor, which selectively blocks the binding of Ang Ⅱ with AT1 receptor, inhibits vasoconstriction and aldosterone release, and produces antihypertensive effect. Irbesartan can also reduce the electrical remodeling of the myocardium, thus reducing the mortality of patients with hypertension. Irbesartan is the most effective drug for the treatment of hypertension and cardiovascular diseases.

 新闻封皮.jpg

Azide is easy to be introduced into irbesartan synthesis process. Azides can inhibit the activities of cytochrome oxidase and many other enzymes, and lead to abnormal phosphorylation and cell respiration. Azide can stimulate respiration, enhance cardiac power, increase hypertension, general spasm, depression and shock in large dose, and it is highly toxic. The half lethal concentration is 37.4mg/kg.


According to the detection requirements of irbesartan in Chinese Pharmacopoeia 2020 edition, anion exchange chromatographic column, conductivity detector and suppressed conductivity detection method were used to analyze the sample solution after online treatment by valve switching online matrix elimination method.

 

1.Instrument and test conditions

1597394658107966.png

CIC-D180 intelligent ion chromatograph


Ion chromatography conditions

Suppressor:SH-A-6

Suppressor current :150 mA

Flow rate: 1.0 ml / min

Temperature of column oven: 35 ℃


Elution conditions of eluent

0-25 10mmol / L

25-32 10mmol / L

32-37 40mmol / L

37-38 40mmol / L

38-45 10mmol / L

45-    102mmol / L


2.Reagent and standard solution

Potassium bromide: reference material 

Sodium azide: reference material

Potassium nitrate: reference material

Ultrapure water: resistivity ≥ 18.25 m Ω· cm (25 ℃)

Acetonitrile: chromatographic purity

Methanol: chromatographic purity


The above test data show that the instrument runs stably and all test data meet the requirements of verification standard.

 1597394714114223.png

 Standard spectrogram


SHINE CIC-D180 intelligent ion chromatograph realizes remote control of mobile app and sharing of information resources, which greatly improves the work efficiency. The modular design also provides more possibilities for analysts with different requirements. It can be used as a dual system and can be equipped with a variety of detectors.